Media Dehydration and Biological Recovery Efficiency
Single Use Device Media Dehydration and Biological Recovery Efficiency

The scope of this study was to verify the performance of collecting contamination from continuous three-hour samplings using the BioCapt® Single-Use manufactured by Particle Measuring Systems under unidirectional airflow, at 22 °C with 48% relative humidity. All four test results aligned with acceptance criteria, and visual inspection of the media post-sampling was satisfactory. With the backing of this study, BioCapt Single-Use can be used for three-hour sampling with no change to its monitoring capabilities.